Chrysalis Intelligence · Sponsors · Wright Medical Technology

Wright Medical Technology

Memphis, TN, US · 1 510(k) clearances · 25 recalls

Total clearances
1
2023-02-17 → 2023-02-17
Last 12 months
0
Product codes
1
Most recent
2023-02-17
Trend

Clearance cadence by quarter

Portfolio

Product codes worked

Code Device name Class Specialty Clearances Most recent
HWC Screw, Fixation, Bone Class II Orthopedic 1 2023-02-17
Safety

Recalls (25)

Class I (serious)
0
Risk of injury / death
Total recalls
25
Currently ongoing
16
Most recent
2024-03-06
DateSeverityStatusRecall # ProductCodeReason
2024-03-06 Class II Ongoing Z-1477-2024 stryker Infinity Resection Adjustment Block, REF 33600030, ankle arthroplasty — Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have …
2023-06-13 Class II Ongoing Z-2105-2023 Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile. — The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventi…
2023-04-10 Class II Ongoing Z-1467-2023 stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A221RT, bone stabilization and fixation — Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.
2023-04-10 Class II Ongoing Z-1466-2023 stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A220RT, bone stabilization and fixation — Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.
2023-02-02 Class II Ongoing Z-1170-2023 ORTHOLOC 3DSi CLAW II 4 HOLE Plate, Size 25MM, REF 40241425, Bone fixation device — Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting…
2023-02-02 Class II Ongoing Z-1169-2023 ORTHOLOC 3DSi CLAW II 4 HOLE Plate, Size 25MM, REF 402S0425, Bone fixation device — Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting…
2022-09-30 Class II Ongoing Z-0140-2023 Stryker INBONE Tibial Tray, Left Plasma Spray Ti6Al4V, CPti, Use w/ 18mm Stem Base, REF 200252904, Size 4 — The tibial tray lock detail is oversized (larger than specification).
2022-09-30 Class II Ongoing Z-0139-2023 Stryker INBONE Tibial Tray, Left Plasma Spray Ti6Al4V, CPti, Use w/ 16mm Stem Base, REF 200252903, Size 3 — The tibial tray lock detail is oversized (larger than specification).
2022-08-03 Class II Ongoing Z-0006-2023 stryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Rep… — The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected…
2022-08-03 Class II Ongoing Z-0005-2023 stryker EVOLVE, Proline Stem, CoCr, Size: +2, O: 705mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repai… — The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected…
2022-07-27 Class II Ongoing Z-1567-2022 Stryker SWANSON Flexspan Finger Joint Implant W/O Grommets Silicone, Size: 00, REF 4700020, Sterile. — An incorrect sizing label was applied to the carton packaging.
2022-06-14 Class II Ongoing Z-1366-2022 Wright EVOLVE TRIAD Plate Cutter, REF 49510120 — The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilizat…
2022-06-02 Class II Ongoing Z-1291-2022 Stryker INFINITY Resect Guide for INBONE Talus, Size 4, Model #33620254, non-sterile, an instrument used with the INFINI… — The instrument is incorrectly color coded.
2022-03-10 Class II Terminated Z-0853-2022 EasyFuse Dynamic Compression System Instrument Pack — The drill and adjustable drill guide can jam/bind intraoperatively resulting in the drill guide breaking.
2022-02-15 Class II Terminated Z-0786-2022 Wright Medical Technology, Inc. ORTHOLOC Non-Locking Screw 2mm x 8mm, REF 5201120008, non-sterile, a component of the OR… — An incorrect screw is packaged in the product, which contains a locking screw instead of a non-locking screw.
2022-02-09 Class II Terminated Z-0679-2022 The DARCO¿ Locking Bone Plate System is designed with rhombus (parallelogram) plates of biocompatible titanium. The pla… — Product was labelled with the incorrect manufacturing and distribution dates.
2021-12-03 Class II Ongoing Z-0401-2022 Wright Medical Technology Inc. CHARLOTTE MTP Hex Screw, Model #41122718, 2.7mm x 18mm, a component of the CHARLOTTE Lisf… — Incorrect product is contained in the packaging.
2021-12-03 Class II Ongoing Z-0393-2022 Stryker DARCO Screw, Locking, Ti6A14V, REF DC2825016, 2.7mm x 16mm. — The incorrect product is contained in the packaging.
2021-12-03 Class II Ongoing Z-0392-2022 Stryker CLAW II ORTHOLOC 3DSi Plate, Hole Qty: 4, 30mm, REF 40240430. — The incorrect product is contained in the packaging.
2021-11-04 Class II Terminated Z-0316-2022 INBONE Total Ankle System, INBONE POLY INSERT SZ 2+ 14mm SULCUS — The package contents and package labeling do not match.
2021-11-04 Class II Terminated Z-0315-2022 INBONE Total Ankle System, INBONE POLY INSERT SZ 2 10mm SULCUS — The package contents and package labeling do not match.
2020-08-26 Class II Terminated Z-2989-2020 EVOLVE STEM 8.5MM Model # 496S085 — Lack of sterility assurance
2020-07-28 Class II Terminated Z-2797-2020 Screw Removal Case Assembly, Part Number 1985BASE - Product Usage: The Wright Medical hardware removal kit is designed t… — Hardware removal kit does not include drivers to remove Dart-Fire 2.5 and 3.0mm Headless screws.
2020-07-28 Class II Terminated Z-2796-2020 Hardware removal kit, Part Number 1985KIT1 - Product Usage: The Wright Medical hardware removal kit is designed to facil… — Hardware removal kit does not include drivers to remove Dart-Fire 2.5 and 3.0mm Headless screws.
2020-04-09 Class II Terminated Z-1924-2020 WRIGHT INBONE Tibial Tray, Size: 2, Plasma Spray, Left, REF 2000252902 - Product Usage: is intended to give a patient li… — One lot of INBONE Tibial Trays is missing the plasma coating.
Filings

All 510(k) clearances (1)

Date K-number Device Code Class Specialty
2023-02-17 K221645 PROstep(TM) MIS 5mm Chamfer Screw System HWC Class II Orthopedic