Chrysalis Intelligence · Sponsors · Terumo Cardiovascular Systems Corporation

Terumo Cardiovascular Systems Corporation

Ann Arbor, MI, US · 3 510(k) clearances · 20 recalls

Total clearances
3
2021-11-18 → 2024-04-25
Last 12 months
0
Product codes
3
Most recent
2024-04-25
Trend

Clearance cadence by quarter

Portfolio

Product codes worked

Code Device name Class Specialty Clearances Most recent
DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass Class II Cardiovascular 1 2024-04-25
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass Class II Cardiovascular 1 2023-04-18
KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type Class II Cardiovascular 1 2021-11-18
Safety

Recalls (20)

Class I (serious)
0
Risk of injury / death
Total recalls
20
Currently ongoing
6
Most recent
2025-05-28
DateSeverityStatusRecall # ProductCodeReason
2025-05-28 Class II Ongoing Z-2018-2025 CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Tri… Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additional…
2025-03-03 Class II Ongoing Z-1492-2025 Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754 During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. …
2024-02-20 Class II Ongoing Z-1401-2024 Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 6… Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential my…
2023-03-01 Class II Ongoing Z-1347-2023 CAPIOX FX25 Hollow Fiber Oxygenator, Catalog Numbers 3CX*FX25REC, 3CX*FX25RWC Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak a…
2023-03-01 Class II Ongoing Z-1346-2023 CAPIOX NX19 Hollow Fiber Oxygenator, Catalog Number 3CX*NX19RE Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak a…
2023-03-01 Class II Ongoing Z-1345-2023 CAPIOX RX15 Hollow Fiber Oxygenator, Catalog Number 3CX*RX15RE30 Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak a…
2022-02-03 Class II Terminated Z-0949-2022 Cardiovascular Procedure Kit catalog # 76645 & 73806 Potential for Cardiovascular Procedure Kit packaging damage that occurred during shipping.
2021-06-08 Class II Terminated Z-2350-2021 Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188 intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The cont…
2021-04-16 Class II Terminated Z-1791-2021 The Sarns TCM II (system) is a source of temperature-controlled water for blood heat exchangers used in an extracorporea… Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol wi…
2021-04-16 Class II Terminated Z-1790-2021 The Sarns Temperature Control and Monitor unit (TCM) is a source of temperature-controlled water for blood heat exchange… Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol wi…
2021-04-16 Class II Terminated Z-1789-2021 The Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directl… Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol wi…
2021-03-02 Class II Terminated Z-1367-2021 Consist of a distal connection mechanism (either standard Terumo Ball Quick Connect or Terumo Shaft Quick Connect), an a… There is a potential for the stainless steel locking plates on the device's sternal retractor locking mechanism to fracture.
2020-11-11 Class II Terminated Z-0565-2021 (1) Capiox FX25 (West) Advance Oxygenator- with left port, 4-liter Reservoir, XC modified to Advance, Sterile Catalog… Capiox FX25 Advance oxygenator may be the incorrect configuration (East vs. West)
2020-08-25 Class II Terminated Z-2961-2020 Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x3/32, Catalog Number 5791 - Product Usage: intended for use… Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped w…
2020-08-25 Class II Terminated Z-2960-2020 Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 1/4x1/16 UAS BLK, Catalog Number 5785 - Product Usage: intended… Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped w…
2020-08-25 Class II Terminated Z-2959-2020 Terumo Advanced Perfusion System 1 Ultrasonic Air Sensor 3/8x3/32, Catalog Number 5773 - Product Usage: intended for use… Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped w…
2020-01-10 Class II Terminated Z-1111-2020 Terumo Advanced Perfusion Assembly Pump Guts, 4 Inch, Catalog Number 801805 - Product Usage: is indicated for use for up… These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a n…
2020-01-10 Class II Terminated Z-1110-2020 Terumo Advanced Perfusion Assembly Pump Guts, 6 Inch, Catalog Number 801798 - is indicated for use for up to 6 hours in … These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a n…
2020-01-10 Class II Terminated Z-1109-2020 Terumo Advanced Perfusion System 1 Roller Pump, 6 Inch, Catalog Number 816571 - Product Usage: is indicated for use for … These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a n…
2020-01-10 Class II Terminated Z-1108-2020 Terumo Advanced Perfusion System 1 Roller Pump, 4 Inch, Catalog Number 816570 - Product Usage: is indicated for use for … These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a n…
Filings

All 510(k) clearances (3)

Date K-number Device Code Class Specialty
2024-04-25 K234065 CDI OneView Monitoring System DRY Class II Cardiovascular
2023-04-18 K221895 Terumo Advanced Perfusion System 1 DTQ Class II Cardiovascular
2021-11-18 K200091 Capiox iCP Centrifugal Pump KFM Class II Cardiovascular