Chrysalis Intelligence · Sponsors · Shimadzu Corporation Medical Systems Division

Shimadzu Corporation Medical Systems Division

Kyoto, JP · 2 510(k) clearances · 6 recalls

Total clearances
2
2022-07-28 → 2024-05-24
Last 12 months
0
Product codes
2
Most recent
2024-05-24
Trend

Clearance cadence by quarter

Portfolio

Product codes worked

Code Device name Class Specialty Clearances Most recent
KPS System, Tomography, Computed, Emission Class II Radiology 1 2024-05-24
OWB 1 2022-07-28
Safety

Recalls (6)

Class I (serious)
0
Risk of injury / death
Total recalls
6
Currently ongoing
6
Most recent
2025-10-16
DateSeverityStatusRecall # ProductCodeReason
2025-10-16 Class II Ongoing Z-0499-2026 Ceiling Type X-Ray Tube Support Models CH-200 and CH-200M used with RADspeed, SonialVision and FLUOROspeed Systems T… Due to an issue (potential breakage) with the shaft for the mounting of the X-ray tube assembly that could potentially result assembly coming off the rotating s…
2024-09-06 Class II Ongoing Z-0317-2025 The FLUOROspeed is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studie… X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography…
2024-09-06 Class II Ongoing Z-0316-2025 SONIALVISION G4 is a multi-purpose X-ray R/F system. The equipment is suitable for radiographic and fluoroscopic examina… X-ray R/F system software with certain versions of the control board may occasionally fail to display images when switching from Fluoroscopy mode to Radiography…
2024-09-06 Class II Ongoing Z-0139-2025 Trinias unity, Digital Angiography System with Catheterization Table, Model: KS-100 This system consists of the followin… If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using dire…
2024-01-29 Class II Ongoing Z-1292-2024 Shimadzu, Trinias, Digital Angiography System Due to capacitor manufacturing issue, their is a potential that Flat Panel Device may have abnormal display or no function.
2023-04-10 Class II Ongoing Z-2270-2023 Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa and G4… Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label b…
Filings

All 510(k) clearances (2)

Date K-number Device Code Class Specialty
2024-05-24 K240827 SET-5002 KPS Class II Radiology
2022-07-28 K221922 Trinias OWB