| 2021-09-10 |
Class II
|
Terminated |
Z-0120-2022 |
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Quantitative Controls (IVD)- For use in monitoring the performance… |
— |
VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Quantitative Controls Outside of Expected Control Range, potential for delayed results |
| 2020-12-15 |
Class II
|
Terminated |
Z-0845-2021 |
VITROS XT Chemistry Products ALB-TP Slides (ALB is not affected) - IVD quantitatively measures total protein (TP) concen… |
— |
Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides |
| 2020-12-15 |
Class II
|
Terminated |
Z-0844-2021 |
VITROS Chemistry Products TP Slides- IVD measure total protein (TP) concentration in serum and plasma using VITROS 250/3… |
— |
Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides |
| 2020-12-11 |
Class II
|
Ongoing |
Z-0770-2021 |
VITROS Chemistry Products CKMB Slides: (1)VITROS CKMB Slides (60 Slide cartridge) Product Code:8058232 (2)VITROS… |
— |
VITROS Chemistry Products CKMB Slides Potential for Falsely Elevated Results Due to CK-MM Interference |
| 2020-10-16 |
Class III
|
Ongoing |
Z-0471-2021 |
VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance… |
— |
Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatu… |
| 2020-09-30 |
Class II
|
Terminated |
Z-0451-2021 |
VITROS¿ Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control - Product Usage: For use in monitoring the performance of … |
— |
Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased si… |
| 2020-09-30 |
Class II
|
Terminated |
Z-0450-2021 |
VITROS Immunodiagnostic Products-SARS-CoV-2 IgG Calibrator - Product Usage: In-Vitro Diagnostic (EUA) an aid in identify… |
— |
Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased si… |
| 2020-07-20 |
Class II
|
Terminated |
Z-2879-2020 |
VITROS XT7600 Integrated System, Product code 6844461 |
— |
When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous re… |
| 2020-07-20 |
Class II
|
Terminated |
Z-2878-2020 |
VITROS XT3400 Chemistry System, Product code 6844458 |
— |
When processing MicroSlides, slides that utilize Slot 18 in the CM rotor may be mis-positioned. As a result, certain VITROS MicroSlides may produce erroneous re… |
| 2020-07-17 |
Class II
|
Terminated |
Z-2771-2020 |
VITROS Anti-SARS-CoV-2 IgG Reagent Pack, COV2IgG, VITROS CoV2G. Catalog number 6199919, UDI 10758750033355 - Product Usa… |
— |
The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality … |
| 2020-07-17 |
Class II
|
Terminated |
Z-2770-2020 |
VITROS Anti-SARS-CoV-2 Total Reagent, COV2TOT, VITROS CoV2T. Catalog number 6199922, UDI 10758750033386 - Product Usage:… |
— |
The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality … |
| 2020-05-20 |
Class II
|
Terminated |
Z-2484-2020 |
VITROS XT 7600 Integrated System Software Versions 3.5.1 and below Product Code:6844461 Unique Identifier: 10758… |
— |
Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances |
| 2020-05-20 |
Class II
|
Terminated |
Z-2483-2020 |
VITROS 5600 Integrated System - Refurbished Software Versions 3.3.3 and below Product Code: 6802915 Unique Identifi… |
— |
Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances |
| 2020-05-20 |
Class II
|
Terminated |
Z-2482-2020 |
VITROS 5600 Integrated System Software Versions 3.3.3 and below Product Code: 6802413 Unique Identifier: 1075875000… |
— |
Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances |
| 2020-05-20 |
Class II
|
Terminated |
Z-2481-2020 |
VITROS 3600 Immunodiagnostic System - Refurbished Product Code: 6802914 Unique Identifier: 10758750007103 |
— |
Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances |
| 2020-05-20 |
Class II
|
Terminated |
Z-2480-2020 |
VITROS 3600 Immunodiagnostic System Software Versions 3.3.3 and below Product Code: 6802783, Unique Identifier: 1075875… |
— |
Modifications to MicroWell Qualitative Assay Parameters Not Retained- may be inadvertently reset to default in certain circumstances |
| 2020-03-10 |
Class III
|
Terminated |
Z-1834-2020 |
VITROS Immunodiagnostic Products TT4 Reagent Pack, Product Code 8744468 |
— |
The specified reagent packs exhibit an increase in the frequency of unsuccessful calibrations. |