| 2024-04-17 |
Class II
|
Ongoing |
Z-1784-2024 |
Intelli-C: Right side suspended Tabletop, REF: 03400000; Left side suspended Tabletop, REF: 03400010, used with the Inte… |
— |
X-ray system C-arm may experience uncontrolled motion if 1) AC motor controller firmware has a faulty setting, 2) the Motor/gear assembly is worn, 3) CAT moveme… |
| 2023-04-27 |
Class II
|
Ongoing |
Z-1669-2023 |
Intelli-C, Left side suspended Tabletop, REF: 03400010 |
— |
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is … |
| 2023-04-27 |
Class II
|
Ongoing |
Z-1668-2023 |
Intelli-C, Right side suspended Tabletop, REF: 03400000 |
— |
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is … |
| 2022-01-21 |
Class II
|
Ongoing |
Z-0526-2022 |
1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusively in combinati… |
— |
Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessor… |