| 2022-04-13 |
Class II
|
Terminated |
Z-1235-2022 |
AUTOLITH Touch 1.9F 375cm Biliary Electrohydraulic Lithotripter Probe |
— |
Product labeled incorrectly. |
| 2021-10-25 |
Class II
|
Ongoing |
Z-0333-2022 |
Boston Scientific AUTOLITH TOUCH Bipolar Electrohydraulic Lithotripter, REF M005466800. The device is designed to be us… |
— |
The languages for Russian, Romanian, Slovak, and Czech have the power settings for high and low switched in the device. |
| 2021-01-15 |
Class II
|
Terminated |
Z-1148-2021 |
Steris CO2MPACT Endoscopic Insufflator system, Catalog Number 72-00310-0 / 710300 - Product Usage: designed to use CO2 a… |
— |
An internal tubing component in the flow path was made with incorrect material which was not tested for use with this device. |
| 2021-01-15 |
Class II
|
Terminated |
Z-1147-2021 |
ConMed 50L abdominal insufflator, Catalog Number 72-00258-0 / GS2000 - Product Usage: The device will maintain an expand… |
— |
An internal tubing component in the flow path was made with incorrect material which was not tested for use with this device. |
| 2020-03-10 |
Class II
|
Terminated |
Z-1885-2020 |
ACMI 9FR PROBE (E4-9F) Manufactured for Olympus Surgical Technologies, America (OSTA); NTI Catalog Number 72-00198-0 - P… |
— |
Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing. |
| 2020-03-10 |
Class II
|
Terminated |
Z-1884-2020 |
AUTOLITH URO-TCH 9FR PROBE 54CM; NTI Catalog Number 9-203-0543 - Product Usage: Probes are used for the fragmentation of… |
— |
Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing. |
| 2020-03-10 |
Class II
|
Terminated |
Z-1883-2020 |
Probe, 9 FR AUTOLITH; NTI Catalog Number 9-900-54 - Product Usage: Probes are used for the fragmentation of urinary/rena… |
— |
As part of an investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing. |