Chrysalis Intelligence · Sponsors · Micro-X Limited

Micro-X Limited

Cloverly Park, AU · 2 510(k) clearances · 10 recalls

Total clearances
2
2020-07-17 → 2021-05-21
Last 12 months
0
Product codes
1
Most recent
2021-05-21
Name variants in FDA filings: Micro-X , Ltd. / Micro-X Limited
Trend

Clearance cadence by quarter

Portfolio

Product codes worked

Code Device name Class Specialty Clearances Most recent
IZL System, X-Ray, Mobile Class II Radiology 2 2021-05-21
Safety

Recalls (10)

Class I (serious)
0
Risk of injury / death
Total recalls
10
Currently ongoing
10
Most recent
2026-03-27
DateSeverityStatusRecall # ProductCodeReason
2026-03-27 Class II Ongoing Z-1918-2026 Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71 X-Ray generator may malfunction resulting in the X-Ray being inoperable.
2025-04-04 Class II Ongoing Z-1670-2025 Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number:MXU-RV35 Product Description: Rover Mobile X… Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
2025-04-04 Class II Ongoing Z-1669-2025 Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71, MXU-RV35 Product Description: Rov… Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
2025-03-28 Class II Ongoing Z-1579-2025 Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile … Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
2025-03-28 Class II Ongoing Z-1578-2025 Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile … Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
2024-10-23 Class II Ongoing Z-0666-2025 MICRO-X Rover Mobile X-ray System, # MXU-RV19 During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the p…
2024-10-22 Class II Ongoing Z-0580-2025 MICRO-X Rover Mobile X-ray System, # MXU-RV35 Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation d…
2023-11-28 Class II Ongoing Z-0665-2025 MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71 Mobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
2023-05-31 Class II Ongoing Z-0664-2025 MICRO-X Rover Mobile X-ray System, # MXU-RV35 Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in…
2022-08-31 Class II Ongoing Z-1786-2022 MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19 Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which…
Filings

All 510(k) clearances (2)

Date K-number Device Code Class Specialty
2021-05-21 K211423 Rover IZL Class II Radiology
2020-07-17 K201488 Rover IZL Class II Radiology