| 2022-07-21 |
Class II
|
Ongoing |
Z-1683-2022 |
Medline CHEST TUBE REORDER NO: DYNJ36762A |
— |
Due to a non-sterile component (tape) being place within the sterile portion of the Chest Tube Convenience Kit |
| 2021-10-27 |
Class II
|
Terminated |
Z-0362-2022 |
Bedside PICC CDS Convenience Kit |
— |
NaCl 250 mL Bag Inj expired 04/2020 while the kit expiration date was after 04/2020. |
| 2021-10-27 |
Class II
|
Terminated |
Z-0361-2022 |
Venous Access Pack-LF Convenience Kit |
— |
NaCl 250 mL Bag Inj expired 04/2020 while the kit expiration date was after 04/2020. |
| 2021-10-27 |
Class II
|
Terminated |
Z-0360-2022 |
Major Vascular CDS Convenience Kit |
— |
NaCl 250 mL Bag Inj expired 04/2020 while the kit expiration date was after 04/2020. |
| 2021-10-27 |
Class II
|
Terminated |
Z-0359-2022 |
Neuro Convenience Kit |
— |
NaCl 250 mL Bag Inj expired 04/2020 while the kit expiration date was after 04/2020. |
| 2021-10-27 |
Class II
|
Terminated |
Z-0358-2022 |
EVLP Convenience Pack/Kit |
— |
NaCl 250 mL Bag Inj expired 04/2020 while the kit expiration date was after 04/2020. |
| 2021-08-06 |
Class II
|
Terminated |
Z-2575-2021 |
NAMIC Stopcocks: (1) NAMIC STOPCOCK, 200 psi (14 bar), REF 70015002, UPN H965700150021, Three-way Stopcock with Rotat… |
— |
Potential for sterile barrier breach. Microscopic pinholes within the sterile pouch could potentially lead to non-sterile conditions. |
| 2021-08-06 |
Class II
|
Terminated |
Z-2574-2021 |
NAMIC Manifolds. (1) NAMIC Angiographic MANIFOLD, Medium Pressure, REF 64038301, UPN H965640383011, 3 Valve Manifo… |
— |
Potential for sterile barrier breach. Microscopic pinholes within the sterile pouch could potentially lead to non-sterile conditions. |
| 2021-07-01 |
Class II
|
Terminated |
Z-2234-2021 |
Medline Arthroscopy Kit, convenience kit, REF DYNJ904304B, containing expired component NaCl 3000mL Irrigation Bag (exp… |
— |
The kits may contain an expired component. |
| 2021-07-01 |
Class II
|
Terminated |
Z-2233-2021 |
Medline Total Knee, convenience kit, REF DYNJ907129B, containing expired component NaCl 3000mL Irrigation Bag (exp. dat… |
— |
The kits may contain an expired component. |
| 2021-07-01 |
Class II
|
Terminated |
Z-2232-2021 |
Medline Knee Arthroscopy Kit, convenience kit, REF DYNJ902838C, containing expired component NaCl 3000mL Irrigation Bag… |
— |
The kits may contain an expired component. |
| 2021-07-01 |
Class II
|
Terminated |
Z-2231-2021 |
Medline Shoulder CDS convenience kits, REF DYNJ902837D and REF DYNJ902837F, containing expired component NaCl 3000mL Irr… |
— |
The kits may contain an expired component. |
| 2021-05-25 |
Class II
|
Terminated |
Z-2080-2021 |
Cotton Rounds/Pads under the following brand names: 1. BEAUTY 360 Extra Large Premium Cotton Ovals - two textures soft… |
— |
Potential for mold contamination (Aspergillus vadensis) |
| 2021-04-30 |
Class II
|
Ongoing |
Z-1841-2021 |
Medline Tissue Recovery Instrument Set, Reorder #DYNJ64347A, packaged 3/case, Sterile, Rx. Convenience kit used to r… |
— |
The devices may not meet the minimum required sterility assurance level. |
| 2021-03-15 |
Class II
|
Ongoing |
Z-1407-2021 |
Diluent, Stat-Site B-HB, 6x 3ML - Product Usage: Intended to be used with the Stat-Site test strips or KetoSite test car… |
— |
Possible false negative or false positive results due to the product being compromised during shipment. |
| 2021-01-25 |
Class II
|
Terminated |
Z-1249-2021 |
RNA MEDICAL, QCHGB HEMOGLOBIN CONTROL, MULTI, REF QC HGB - Product Usage: monitoring the performance of Hemocue Hemoglob… |
— |
Product was compromised during shipment. |
| 2021-01-15 |
Class III
|
Terminated |
Z-1135-2021 |
S-O303000 CONTROL, STAT-SITE B-HB, 2X3ML,1LOW,1HIGH - Product Usage: intended for the quantitative determination of the … |
— |
Medline Industries, Inc. is removing specific lots of STAT-Site¿ Controls (Low and High) S-O303000, from the market, due to the product being compromised durin… |
| 2020-11-13 |
Class II
|
Terminated |
Z-0783-2021 |
Cardiac Pack - LF, Model DYNJ31903I - Product Usage: This is a type of convenience kit used in various surgeries includi… |
— |
The cylindrical sponge component is not x-ray detectable. |
| 2020-11-13 |
Class II
|
Terminated |
Z-0782-2021 |
Greffe Aorte-LF (product info is in French); Model DYNJ31877F - Product Usage: This is a type of convenience kit used in… |
— |
The cylindrical sponge component is not x-ray detectable. |
| 2020-11-13 |
Class II
|
Terminated |
Z-0781-2021 |
Thyroid Pack CKHA-LF, Model DYNJ50352B - Product Usage: This is a type of convenience kit used in various surgeries incl… |
— |
The cylindrical sponge component is not x-ray detectable. |
| 2020-11-13 |
Class II
|
Terminated |
Z-0780-2021 |
Simple Neck Dissection, Model DYNJ47481C - Product Usage: This is a type of convenience kit used in various surgeries in… |
— |
The cylindrical sponge component is not x-ray detectable. |
| 2020-11-13 |
Class II
|
Terminated |
Z-0779-2021 |
Resect. Intest. Laparo-LF, Model DYNJ42667A - Product Usage: This is a type of convenience kit used in various surgeries… |
— |
The cylindrical sponge component is not x-ray detectable. |
| 2020-11-05 |
Class II
|
Terminated |
Z-0615-2021 |
RevMedX Trauma Dressings are packaged in a Tyvek Peel pouch, 1 Trauma Dressing per Tyvec Peel Pouch, 30 Pouches per Case… |
— |
RevMedX Trauma Dressing is being recalled due to package seal integrity. |
| 2020-10-01 |
Class II
|
Terminated |
Z-0297-2021 |
The Quidel Triage BNP Calibrators |
— |
Product compromised during shipment. |
| 2020-09-15 |
Class II
|
Ongoing |
Z-0055-2021 |
Quick Release Gait Belt, 60 |
— |
The buckle on the Gait Belt may break due to degradation of the material due to exposure to gamma radiation. |
| 2020-09-14 |
Class II
|
Terminated |
Z-0068-2021 |
Kit Model #65193091, MTO Left Heart Kit - Product Usage: NAMIC Cath Lab and Interventional Radiology (IR) Kits consist o… |
— |
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities. |
| 2020-09-14 |
Class II
|
Terminated |
Z-0067-2021 |
Kit Model #65220955, MTO Left Heart Kit VA Med San Francisco - Product Usage: NAMIC Cath Lab and Interventional Radiolog… |
— |
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities. |
| 2020-09-14 |
Class II
|
Terminated |
Z-0066-2021 |
Kit Model #650600111, MTO Left Heart St Francis Hosp - PG - Product Usage: NAMIC Cath Lab and Interventional Radiology (… |
— |
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities. |
| 2020-09-14 |
Class II
|
Terminated |
Z-0065-2021 |
Kit Model #650305414, MTO Left Heart with DT Boulder Com Hosp - Product Usage: NAMIC Cath Lab and Interventional Radiolo… |
— |
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities. |
| 2020-09-14 |
Class II
|
Terminated |
Z-0064-2021 |
Kit Model #650304917, MTO Protection Station Plus with Drape - Product Usage: NAMIC Cath Lab and Interventional Radiolog… |
— |
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities. |
| 2020-09-14 |
Class II
|
Terminated |
Z-0063-2021 |
Kit Model #650301023, MTO Radiology Kit Cleveland Clinic - Product Usage: NAMIC Cath Lab and Interventional Radiology (I… |
— |
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities. |
| 2020-09-14 |
Class II
|
Terminated |
Z-0062-2021 |
Kit Model #650301018, MTO Vascular Surgery - Cleveland Clinic PG - Product Usage: NAMIC Cath Lab and Interventional Radi… |
— |
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities. |
| 2020-09-14 |
Class II
|
Terminated |
Z-0061-2021 |
Kit Model #65021652, MTO Left Heart Kit United Reg Health - Product Usage: NAMIC Cath Lab and Interventional Radiology (… |
— |
Non-sterile kits intended to be sterilized prior to use were sold to accounts without sterilization capabilities. |
| 2020-09-01 |
Class II
|
Terminated |
Z-0058-2021 |
ClearPro Closed Suction Catheter, Model DYNCPDS14T, 14 Fr FOR ENDOTRACHEAL TUBE - Product Usage: Catheters are intended… |
— |
There have been reports of the suction catheter coming apart from the device assembly during use. |
| 2020-08-12 |
Class II
|
Terminated |
Z-3013-2020 |
Central Line Dressing Change Tray |
— |
Potential of incomplete seals impacting the sterility of the kit. |
| 2020-07-28 |
Class II
|
Terminated |
Z-3054-2020 |
Control, Calibrator Verified, Triage BNP, Model ALR88755 |
— |
Product was shipped with refrigeration (ice) instead of frozen (dry ice). |
| 2020-07-09 |
Class II
|
Terminated |
Z-2766-2020 |
Endoscopic Cleaning Brush - Product Usage: Endoscopic cleaning brushes are single use devices intended to clean the inte… |
— |
Potential risk of brush detachment during cleaning. |
| 2020-06-30 |
Class II
|
Ongoing |
Z-2725-2020 |
ClearPro Trach T-Piece Closed Suction Catheter, Model DYNCPTP14T, 14 Fr 12.5 IN FOR TRACHEOSTOMY TUBE - Product Usage: C… |
— |
There are reports of the suction catheter coming apart from the device assembly during use. |
| 2020-06-30 |
Class II
|
Ongoing |
Z-2724-2020 |
ClearPro T-Piece Closed Suction Catheter, Model DYNCPTP14, 14 Fr 21.3 IN FOR ENDOTRACHEAL TUBE - Product Usage: Catheter… |
— |
There are reports of the suction catheter coming apart from the device assembly during use. |
| 2020-03-09 |
Class II
|
Terminated |
Z-2040-2020 |
Custom convenience kits containing Medline brand Sterile Pre-Saturated Povidone Iodine Sponge Products, labeled as follo… |
— |
Medline identified that there may be a potential for voids in the packaging seal and loss of sterility on the outside of the foil pouch. |
| 2020-02-10 |
Class II
|
Terminated |
Z-1546-2020 |
Medline Sterile 3mL Syringes with 21GA x 1 Inch Needle 80 each REF DYND3ML21G1A lot 18JBH655 (01) 1088438982153. |
— |
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging. |
| 2020-02-10 |
Class II
|
Terminated |
Z-1545-2020 |
Medline Winged Sponges Presaturated W/PVP REF DYND70289 DYND70289H Expiration 06/30/2022. |
— |
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging. |
| 2020-02-10 |
Class II
|
Terminated |
Z-1544-2020 |
Medline Stick Sponge Presaturated W/PVP REF DYND70288 DYND70288H 2/PK Expiration 06/30/2022. |
— |
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging. |
| 2020-01-21 |
Class II
|
Terminated |
Z-1358-2020 |
Medline ROBOTIC-LF REF CDS984543B Lot 20AKA927 Expiration 2020-09-30 GTIN (01)10888277056350 |
— |
The sterile pack inside the Complete Delivery System (CDS) did not go through a complete sterilization cycle, therefore the sterility of these components cannot… |
| 2020-01-21 |
Class II
|
Terminated |
Z-1357-2020 |
Medline HOI LUMBAR REF DYNJ902823G Lot 20AKA895 Expiration 2021-05-31 1 each GTIN (01)10193489219180 |
— |
The sterile pack inside the Complete Delivery System (CDS) did not go through a complete sterilization cycle, therefore the sterility of these components cannot… |
| 2020-01-07 |
Class II
|
Terminated |
Z-0865-2020 |
Medline DYNDA1928 VP TRAY WITH PVP Packaged in Mexico GLOVE:Made in Malaysia APPLICATOR:Made in China STERILE (01)… |
— |
There is potential for the sterile wrap to become loose while in transit and compromise the sterile barrier. |