| 2026-02-10 |
Class II
|
Ongoing |
Z-1574-2026 |
Artegraft Vascular Graft; REF#: AG740; |
— |
Labeling mix-up resulting in the incorrect lot outer packaging of product. |
| 2025-09-25 |
Class II
|
Ongoing |
Z-0289-2026 |
Artegraft¿ Collagen Vascular Graft; REF: AG1015; |
— |
The device was incorrectly packed in the wrong size labeled outer packaging. |
| 2025-09-02 |
Class II
|
Ongoing |
Z-0034-2026 |
Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 71… |
— |
The bovine carotid arteries used to produce the impacted grafts from affected lots were sourced using raw material from a supplier that had not been reviewed an… |
| 2025-08-25 |
Class II
|
Ongoing |
Z-0072-2026 |
Artegraft Collagen Vascular Graft. Model/Catalog/Part Numbers: AG630M, AG636M, AG730M, AG740M, AG845M. The Artegr… |
— |
Devices were distributed with incorrect label without CE and UKCA mark, missing patient leaflet and patient implant card. |
| 2025-04-11 |
Class II
|
Ongoing |
Z-1752-2025 |
Pruitt Irrigation Occlusion Catheter. Catalog Numbers: 2102-09, E2102-09. |
— |
During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier). |
| 2025-04-11 |
Class II
|
Ongoing |
Z-1751-2025 |
Pruitt Occlusion Catheter. Catalog Numbers: 2103-36, 2103-46, 2103-56, E2103-36, E2103-46, E2103-56 |
— |
During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier). |
| 2025-04-11 |
Class II
|
Ongoing |
Z-1750-2025 |
TufTex Over-the-Wire Embolectomy Catheter. All Catalog numbers: 1651-34, 1651-38, 1651-44, 1651-48, 1651-58, 1651-64, 1… |
— |
During internal product testing, it was observed that some packages to have incomplete seals (sterile barrier). |
| 2024-04-17 |
Class II
|
Ongoing |
Z-1824-2024 |
Syntel Silicone Thrombectomy Catheter, Sterile, Rx Only. Used in vascular grafts and peripheral venous thrombectomies. |
— |
The guide tip can become damaged and result in the tip detaching. |
| 2023-07-13 |
Class II
|
Ongoing |
Z-2468-2023 |
The Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be used as grafts in the in situ … |
— |
Burrs (molding flash) on the cutter may tear the harvested vien |
| 2022-09-30 |
Class II
|
Ongoing |
Z-0228-2023 |
TufTex Embolectomy Catheter, Model No. This catheter has a latex balloon and a single lumen extrusion. It is inserte… |
— |
Three lots tied by two trainees were released without performing the required pull test. If tied with inadequate strength, the proximal ligature could slip towa… |
| 2021-04-06 |
Class II
|
Terminated |
Z-1529-2021 |
LeverEdge Contrast Injector: 35 cc syringe with ISO 594 female luer lock - Product Usage: intended to be used to perfuse… |
— |
The firm received a complaint regarding particulate found in the packaging foam that could contaminate the sterile field. |
| 2020-03-23 |
Class I
|
Ongoing |
Z-1567-2020 |
LeMaitre 5F Plus Over the Wire Embolectomy Catheter, Model Nos. 1651-84, e1651-84, 1651-88, e1651-88. The LeMaitre Ov… |
— |
There has been an increased trend in reports of catheters failing to deflate during use, which may ultimately lead to tip separation. |