| 2026-01-20 |
Class II
|
Ongoing |
Z-1543-2026 |
GEM Premier 5000; Part No. 00055445010 & 00055445011. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1542-2026 |
GEM Premier 5000; Part No. 00055445008. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1541-2026 |
GEM Premier 5000; Part No. 00055445004. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1540-2026 |
GEM Premier 5000; Part No. 00055430011. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1539-2026 |
GEM Premier 5000; Model No. 00055430010. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1538-2026 |
GEM Premier 5000; Part No. 00055430008. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1537-2026 |
GEM Premier 5000; Part No. 00055430004. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1536-2026 |
GEM Premier 5000; Part No: 00055415011. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1535-2026 |
GEM Premier 5000; Part No. 00055415010. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1534-2026 |
GEM Premier 5000; Part No. 00055415008. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1533-2026 |
GEM Premier 5000; Part No. 00055415005. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1532-2026 |
GEM Premier 5000 PAK; Part No. 00055415004. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1531-2026 |
GEM Premier 5000 PAK; Part No. 00055407511. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1530-2026 |
GEM Premier 5000 PAK; Part No. 00055407510. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1529-2026 |
GEM Premier 5000 PAK; Part No. 00055407508. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1528-2026 |
GEM Premier 5000 PAK; Part No. 00055407504. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1527-2026 |
GEM Premier 5000 PAK; Part No. 00055360011. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1526-2026 |
GEM Premier 5000 PAK; Part No. 00055360010. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1525-2026 |
GEM Premier 5000 PAK; Part No. 00055360008. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2026-01-20 |
Class II
|
Ongoing |
Z-1524-2026 |
GEM Premier 5000 PAK, Part No. 00055360004. |
— |
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution … |
| 2025-12-11 |
Class II
|
Ongoing |
Z-1095-2026 |
The product is used for the evaluation of the intrinsic coagulation pathway, APTT substitution test and the monitoring o… |
— |
Potential for microbial contamination. |
| 2025-11-18 |
Class III
|
Ongoing |
Z-0938-2026 |
HemosIL SynthAFax. Partial Thromboplastin Time Tests. |
— |
Recalled lots were manufactured with double the amount of preservative concentration. |
| 2025-02-03 |
Class II
|
Ongoing |
Z-2323-2025 |
HemosIL LMW Heparin Controls; Part Number: 0020300200; |
— |
Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert ac… |
| 2024-12-18 |
Class II
|
Ongoing |
Z-1020-2025 |
HemosIL Heparin Calibrators; Part Number: 0020300600; in vitro diagnostic |
— |
Control indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance ran… |
| 2023-09-29 |
Class II
|
Ongoing |
Z-0359-2024 |
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Glucose and Lactate, Part Numbers 00024315009, 00024330009, 00… |
— |
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 le… |
| 2023-09-29 |
Class II
|
Ongoing |
Z-0358-2024 |
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 0002434500… |
— |
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 le… |
| 2023-09-29 |
Class II
|
Ongoing |
Z-0357-2024 |
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004 |
— |
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 le… |
| 2023-07-13 |
Class II
|
Ongoing |
Z-2578-2023 |
ACL TOP Family 50 Series Models as follows: In vitro diagnostic clinical use in the hemostasis laboratory for coagulati… |
— |
Sample misidentification could occur under specific conditions and patient management altered based on an incorrectly assigned result |
| 2023-02-03 |
Class II
|
Ongoing |
Z-1307-2023 |
ACL ELITE PRO Refurbished)-Analyzer designed specifically for clinical use in the hemostasis laboratory, for coagulation… |
— |
Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments |
| 2023-02-03 |
Class II
|
Ongoing |
Z-1306-2023 |
ACL ELITE Refurbished-Analyzer designed specifically for clinical use in the hemostasis laboratory, for coagulation and/… |
— |
Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments |
| 2023-02-03 |
Class II
|
Ongoing |
Z-1305-2023 |
ACL ELITE PRO-Analyzer designed specifically for clinical use in the hemostasis laboratory, for coagulation and/or fibri… |
— |
Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments |
| 2023-02-03 |
Class II
|
Ongoing |
Z-1304-2023 |
ACL ELITE- Analyzer designed specifically for clinical use in the hemostasis laboratory, for coagulation and/or fibrinol… |
— |
Removes the test definition for HemosIL Liquid Anti-Xa (PN 0020302600 and 0020302601) from the ACL Elite/Elite Pro instruments |
| 2022-04-08 |
Class II
|
Ongoing |
Z-1052-2022 |
HemosIL ReadiPlasTin, Part No. 0020301400 |
— |
Variable and out of specification QC results with HemosIL ReadiPlasTin. IL will remove all lots from the market and convert customers to an alternative prothro… |
| 2022-02-22 |
Class II
|
Ongoing |
Z-1014-2022 |
ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL TOP 300 CTS … |
— |
Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or r… |
| 2021-10-13 |
Class II
|
Ongoing |
Z-0276-2022 |
HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent. |
— |
The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonge… |
| 2021-08-25 |
Class II
|
Ongoing |
Z-0151-2022 |
HemosIL Liquid Anti-Xa- Automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) a… |
— |
Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days |
| 2021-08-25 |
Class II
|
Ongoing |
Z-0150-2022 |
HemosIL Liquid Anti-Xa- automated chromogenic assay for in vitro diagnostic use by laboratory professionals in clinical… |
— |
Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days |
| 2021-08-25 |
Class II
|
Ongoing |
Z-0149-2022 |
HemosIL Liquid Anti-Xa- Automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) a… |
— |
Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days |