| 2024-03-22 |
Class II
|
Ongoing |
Z-1619-2024 |
48" Lead Wires, Replacement Part Number: 1067724-4 |
— |
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to… |
| 2024-03-22 |
Class II
|
Ongoing |
Z-1618-2024 |
20" Lead Wires, Replacement Part Number: 1067724-2 |
— |
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to… |
| 2024-03-22 |
Class II
|
Ongoing |
Z-1617-2024 |
Biomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an established nonunion acqui… |
— |
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to… |
| 2024-03-22 |
Class II
|
Ongoing |
Z-1616-2024 |
Biomet SpinalPak Non-invasive Spine Fusion Stimulator System- Indicated for the treatment of an established nonunion ac… |
— |
Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to… |