Chrysalis Intelligence · Sponsors · Diagnostica Stago, Inc.

Diagnostica Stago, Inc.

Parsippany, NJ, US · 1 510(k) clearances · 13 recalls

Total clearances
1
2026-04-27 → 2026-04-27
Last 12 months
1
Product codes
1
Most recent
2026-04-27
Trend

Clearance cadence by quarter

Portfolio

Product codes worked

Code Device name Class Specialty Clearances Most recent
JPA 1 2026-04-27
Safety

Recalls (13)

Class I (serious)
0
Risk of injury / death
Total recalls
13
Currently ongoing
12
Most recent
2026-02-05
DateSeverityStatusRecall # ProductCodeReason
2026-02-05 Class II Ongoing Z-1585-2026 Brand Name: STA Liatest D-Di Product Name: STA Liatest D-Di Model/Catalog Number: REF 00515 Product Description: The … After receiving customer complaints, investigations confirmed the presence of a positive bias in D-Dimer measurements throughout the entire analytical range.
2026-02-04 Class II Ongoing Z-1491-2026 Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Pro… The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.
2025-12-15 Class II Ongoing Z-1084-2026 Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection i… Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.
2025-09-12 Class II Ongoing Z-0205-2026 Product: STA Liatest D-Di; REF: 00515; A potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.
2025-07-21 Class II Ongoing Z-2476-2025 STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers o… Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
2025-07-21 Class II Ongoing Z-2475-2025 STA-Liquid Anti-Xa 8. Model/Catalog Number: 00322US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, … Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
2025-07-21 Class II Ongoing Z-2474-2025 STA-Liquid Anti-Xa 4. Model/Catalog Number: 00311US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, S… Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
2025-07-21 Class II Ongoing Z-2473-2025 STA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop a… Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
2025-05-21 Class II Ongoing Z-1993-2025 STA - Owren Koller. Model/Catalog Number: 00360 STA - Owren-Koller is a buffer solution intended for use as a diluent… Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%
2023-03-22 Class II Ongoing Z-1577-2023 STA Compact Max fully automatic clinical analyzer, Reference Numbers 58602 and 58612 An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased freque…
2023-03-22 Class II Ongoing Z-1576-2023 STA Compact Automated Multi-Parametric Analyzer, Reference Numbers 58602 and 58612 An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased freque…
2023-03-22 Class II Ongoing Z-1575-2023 STA R Max fully automatic clinical instrument designed to perform tests on human plasmas, Reference Numbers 59013 and 59… An internal investigation identified a bug in this firmware version, resulting in the following:Intermittent shortened coagulation times and an increased freque…
2020-03-04 Class II Terminated Z-1699-2020 STA - Deficient VIII, Product Code: 00725. STA - Deficient VIII is an immuno-depleted human plasma intended for use in … A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause facto…
Filings

All 510(k) clearances (1)

Date K-number Device Code Class Specialty
2026-04-27 K253658 STA Satellite Max® JPA