| 2026-05-27 |
Class II
|
Ongoing |
Z-2574-2026 |
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack … |
— |
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were late… |
| 2026-04-14 |
Class II
|
Ongoing |
Z-2165-2026 |
Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Numbe… |
— |
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend … |
| 2026-04-14 |
Class II
|
Ongoing |
Z-2164-2026 |
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Numbe… |
— |
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend … |
| 2026-04-14 |
Class II
|
Ongoing |
Z-2163-2026 |
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: S… |
— |
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend … |
| 2025-08-28 |
Class I
|
Ongoing |
Z-0318-2026 |
Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model… |
— |
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glu… |
| 2025-07-24 |
Class I
|
Ongoing |
Z-2450-2025 |
Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Numbe… |
— |
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier)… |
| 2025-07-24 |
Class I
|
Ongoing |
Z-2449-2025 |
Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ Android CGM App Model/Catalog N… |
— |
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier)… |
| 2025-07-24 |
Class I
|
Ongoing |
Z-2448-2025 |
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Numbe… |
— |
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier)… |
| 2025-07-24 |
Class I
|
Ongoing |
Z-2447-2025 |
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: S… |
— |
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier)… |
| 2025-07-24 |
Class I
|
Ongoing |
Z-2446-2025 |
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Numbe… |
— |
The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier)… |
| 2025-05-12 |
Class I
|
Ongoing |
Z-1946-2025 |
Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), … |
— |
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or h… |
| 2025-05-12 |
Class I
|
Ongoing |
Z-1945-2025 |
Dexcom One+ Continuous Glucose Monitoring System, SKU(Part Number): STK-D7-013 (MT28285-1), STK-D7-014 (MT28285-1), STK-… |
— |
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or h… |
| 2025-05-12 |
Class I
|
Ongoing |
Z-1944-2025 |
Dexcom G7 Glucose Receiver, SKUs (part number): STK-AT-011(MT26403-0), STK-AT-012 (MT26403-0), STK-AT-013 (MT26403-0),… |
— |
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or h… |
| 2025-05-12 |
Class I
|
Ongoing |
Z-1943-2025 |
Dexcom G6 Glucose Receiver, SKUs (Part Number): STK-FM-001 (MT27408-1) & STK-FR-001 (MT27408-1) |
— |
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or h… |
| 2025-05-05 |
Class II
|
Ongoing |
Z-1906-2025 |
Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) syste… |
— |
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show … |
| 2025-01-27 |
Class II
|
Ongoing |
Z-1204-2025 |
Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OE-001 |
— |
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different… |
| 2025-01-27 |
Class II
|
Ongoing |
Z-1203-2025 |
Dexcom G6, Continuous Glucose Monitoring System, REF: STK-OR-001 |
— |
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different… |
| 2025-01-27 |
Class II
|
Ongoing |
Z-1202-2025 |
Dexcom G6, Continuous Glucose Monitoring System, REFs: STK-OM-001, STR-OM-001, STR-OR-001 |
— |
Under very rare situations, the Dexcom G6 touchscreen receiver may not provide high or low glucose alarms/alerts as designed, which can result in two different… |
| 2020-02-03 |
Class II
|
Terminated |
Z-1562-2020 |
Dexcom G6 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a gluco… |
— |
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous G… |
| 2020-02-03 |
Class II
|
Terminated |
Z-1561-2020 |
Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a gluco… |
— |
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous G… |
| 2020-02-03 |
Class II
|
Terminated |
Z-1560-2020 |
Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System i… |
— |
The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous G… |