| 2020-08-31 |
Class II
|
Terminated |
Z-0079-2021 |
CERTAIN BELLATEK TITANIUM ABUTMENT 6.0MM Reference Number: IEDAT6 |
— |
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may ca… |
| 2020-08-31 |
Class II
|
Terminated |
Z-0078-2021 |
CERTAIN BELLATEK TITANIUM ABUTMENT 5.0MM Reference Number: IEDAT5 |
— |
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may ca… |
| 2020-08-31 |
Class II
|
Terminated |
Z-0077-2021 |
CERTAIN BELLATEK TITANIUM ABUTMENT 4.1MM Reference Number: IEDAT4 |
— |
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may ca… |
| 2020-08-31 |
Class II
|
Terminated |
Z-0076-2021 |
CERTAIN BELLATEK TIN ABUTMENT 5.0MM Reference Number: IEDAN5 |
— |
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may ca… |
| 2020-08-31 |
Class II
|
Terminated |
Z-0075-2021 |
CERTAIN BELLATEK TIN ABUTMENT 4.1MM Reference Number: IEDAN4 |
— |
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may ca… |
| 2020-08-31 |
Class II
|
Terminated |
Z-0074-2021 |
BELLATEK TITANIUM ABUTMENT 4.1MM Reference Number: EDAT4 |
— |
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may ca… |
| 2020-08-31 |
Class II
|
Terminated |
Z-0073-2021 |
BellaTek Encode Defin3DPMSERitive Abutments-3D Printed Model Reference Numbers: 3DPMSER, 3DPMSERQ |
— |
Hex of the BellaTek Encode Definitive Abutments was misaligned by 30 degrees during manufacturing as compared with the provided patient scan data (Images may ca… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1388-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 7.5MM(P) X 4MM(H) Reference Number: IEHA574 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1387-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 6MM(P) X 6MM(H) Reference Number: IEHA566 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1386-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 6MM(P) X 4MM(H) Reference Number: IEHA564 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1385-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 6MM(P) X 4MM(H) Reference Number: IEHA564 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1384-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 6MM(P) X 3MM(H) Reference Number: IEHA563 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1383-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 5.6MM(P) X 4MM(H) Reference Number: IEHA554 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1382-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 5MM(D) X 5.6MM(P) X 3MM(H) Reference Number: IEHA553 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1381-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 7.5MM(P) X 4MM(H) Reference Number: IEHA474 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1380-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 6MM(P) X 6MM(H) Reference Number: IEHA466 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1379-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 6MM(P) X 4MM(H) Reference Number: IEHA464 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1378-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 6MM(P) X 3MM(H) Reference Number: IEHA463 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1377-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 5MM(P) X 6MM(H) Reference Number: IEHA456 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1376-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 5MM(P) X 4MM(H) Reference Number: IEHA454 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1375-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 5MM(P) X 3MM(H) Reference Number: IEHA453 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1374-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 4.1MM(P) X 6MM(H) Reference Number: IEHA446 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1373-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 4.1MM(P) X 4MM(H) Reference Number: IEHA444 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1372-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 4.1MM(D) X 4.1MM(P) X 3MM(H) Reference Number: IEHA443 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1371-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 5MM(P) X 6MM(H) Reference Number: IEHA356 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1370-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 5MM(P) X 4MM(H) Reference Number: IEHA354 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1369-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 5MM(P) X 3MM(H) Reference Number: IEHA353 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1368-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 3.8MM(P) X 6MM(H) Reference Number: IEHA346 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1367-2020 |
CERTAIN¿ BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 3.8MM(P) X 4MM(H) Reference Number: IEHA344 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |
| 2020-01-10 |
Class II
|
Terminated |
Z-1366-2020 |
CERTAIN BELLATEK ENCODE HEALING ABUTMENT 3.4MM(D) X 3.8MM(P) X 3MM(H) Reference Number: IEHA343 |
— |
Resulting impression or scanned data produced from the Healing Abutment lots will result in an incorrect rotation of approximately 30 degrees and/or margin cont… |