| 2026-06-04 |
Class II
|
Ongoing |
Z-2715-2026 |
VITEK 2 AST-N447 TEST KIT Product reference 424620 |
— |
Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures. |
| 2026-01-05 |
Class II
|
Ongoing |
Z-1236-2026 |
VITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem… |
— |
Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or… |
| 2025-06-26 |
Class II
|
Ongoing |
Z-2671-2025 |
VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system c… |
— |
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific… |
| 2024-03-14 |
Class I
|
Ongoing |
Z-1464-2024 |
VITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; … |
— |
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible resu… |
| 2023-02-06 |
Class II
|
Ongoing |
Z-1201-2023 |
MYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects microbiology laboratory instruments … |
— |
For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always be… |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0876-2023 |
PPM LOMBARD PRODUCTS T PHOSPHATE BUFFERSTERILE 20X9ML, CATALOG T8035 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0875-2023 |
PPM LOMBARD PRODUCTS IR&DB - TSA (32ML) 10PLT, CATALOG M1205-IR |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0874-2023 |
PPM LOMBARD PRODUCTS R2A MEDIUM 10PLT, CATALOG M1065 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0873-2023 |
PPM LOMBARD PRODUCTS TSA 10PLT, CATALOG M1040 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0872-2023 |
PPM LOMBARD PRODUCTS CT&TI-TSA + L & P80 10PLT, CATALOG C6045-TI |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0871-2023 |
PPM LOMBARD PRODUCTS CT&IR&DB-TSA + L & P80 10PLT, CATALOG C6045-IR |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0870-2023 |
PPM INDUSTRY COMBOURG BACARA 20PLATES 90MM, CATALOG AEB520100 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0869-2023 |
DCOUNT SYSTEMS REAGENTS CHEMCHROME V14 (6X1.8ML), CATALOG 306-R1026-01 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0868-2023 |
VIDAS CLINICAL VIDAS BRAHMS PROCALCITONIN 60T, CATALOG 30450-01 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0867-2023 |
VIDAS CLINICAL VIDAS ESTRADIOL II 60 TESTS, CATALOG 30431-01 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0866-2023 |
VIDAS CLINICAL VIDAS PROGESTERONE 60 TESTS, CATALOG 30409-01 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0865-2023 |
VIDAS CLINICAL VIDAS LH 60 TESTS, CATALOG 30406-01 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0864-2023 |
VIDAS CLINICAL VIDAS HCG 60 TESTS, CATALOG 30405-01 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0863-2023 |
ETEST CLINICAL COLISTIN CO 256 US B100, CATALOG 537348 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0862-2023 |
ASTUTE REAGENT NEPHROCHECK LIQUID CONTROLS KIT US, CATALOG 500013 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0861-2023 |
VITEK 2 REAGENT AST-XN15 TEST KIT 20 CARDS, CATALOG 423829 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0860-2023 |
ETEST CLINICAL ETEST VANCOMYCIN VA M100 US, CATALOG 423788 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0859-2023 |
ETEST CLINICAL ETEST MEROPENEM MP M100 US, CATALOG 423786 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0858-2023 |
PPM LOMBARD PRODUCTS TSA 3P W NEUTRALIZERS 100 PLT, CATALOG 423723 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0857-2023 |
VITEK 2 REAGENT AST-XN09 TEST KIT 20 CARDS, CATALOG 423425 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0856-2023 |
VITEK 2 REAGENT AST-N801 TEST KIT 20 CARDS, CATALOG 423416 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0855-2023 |
GENE-UP REAGENT GENE-UP Salmonella 2, CATALOG 423105 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0854-2023 |
VITEK 2 REAGENT AST-GN99 TEST KIT 20 CARDS, CATALOG 423102 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0853-2023 |
VITEK 2 REAGENT AST-GN95 TEST KIT 20 CARDS, CATALOG 421982 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0852-2023 |
ETEST CLINICAL ETEST MEROPENEM/VABORBACTAM MEV US S30, CATALOG 421560 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0851-2023 |
ETEST CLINICAL ERAVACYCLINE ERV US S30, CATALOG 421553 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0850-2023 |
VITEK 2 REAGENT AST-GP78 TEST KIT 20 CARDS, CATALOG 421051 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0849-2023 |
ETEST CLINICAL IMIPENEM RELEBACTAM IPR US S30, CATALOG 420927 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0848-2023 |
VITEK 2 REAGENT AST-ST02 TEST KIT 20 CARDS, CATALOG 420915 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0847-2023 |
PPM CLINICAL CHROMID STREPTO B US 20 PLT, CATALOG 419751 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0846-2023 |
ETEST CLINICAL CEFTAZIDIM/AVIBACTAM US S30, CATALOG 419046 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0845-2023 |
PPM LOMBARD PRODUCTS TSA 3P W NEUTRALIZERS 100 PLT, CATALOG 419014 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0844-2023 |
PPM LOMBARD PRODUCTS IRR COUNT-TACT 3P AGAR 100 PLT, CATALOG 418049 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0843-2023 |
VITEK 2 REAGENT AST-GP75 TEST KIT 20 CARDS, CATALOG 415670 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0842-2023 |
PPM CLINICAL CHROMID MRSA S AUREUS 20PLT US, CATALOG 414524 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0841-2023 |
ETEST CLINICAL CEFTOLOZANE/TAZO C/T US S30, CATALOG 414445 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0840-2023 |
GENE-UP REAGENT GENE-UP LYSIS KIT, CATALOG 414057 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0839-2023 |
PPM INDUSTRY COMBOURG CHROMID CARBA US 20 PLT, CATALOG 414012 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0838-2023 |
VITEK 2 REAGENT AST-GN81 TEST KIT 20 CARDS, CATALOG 413438 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0837-2023 |
VITEK 2 REAGENT AST-GN80 TEST KIT 20 CARDS, CATALOG 413437 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0836-2023 |
VITEK 2 REAGENT AST-GN79 TEST KIT 20 CARDS, CATALOG 413436 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0835-2023 |
VITEK 2 REAGENT AST-GN73 TEST KIT 20 CARDS, CATALOG 413404 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0834-2023 |
VITEK 2 REAGENT AST-GN70 TEST KIT 20 CARDS, CATALOG 413401 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0833-2023 |
VITEK 2 REAGENT AST-GN69 TEST KIT 20 CARDS, CATALOG 413400 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0832-2023 |
BIOBALL K.RHIZOPHILA (30X20), CATALOG 412690 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0831-2023 |
ETEST CLINICAL VANCOMYCIN VA 256 US S30, CATALOG 412486 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0830-2023 |
ETEST CLINICAL TOBRAMYCIN TM 256 US S30, CATALOG 412478 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0829-2023 |
ETEST CLINICAL RIFAMPICIN RI 32 US S30, CATALOG 412449 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0828-2023 |
ETEST CLINICAL MINOCYCLINE MC 256 US S30, CATALOG 412407 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0827-2023 |
ETEST CLINICAL MEROPENEM MP 32 US S30, CATALOG 412401 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0826-2023 |
ETEST CLINICAL LEVOFLOXACIN LE 32 US S30, CATALOG 412392 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0825-2023 |
ETEST CLINICAL IMIPENEM IP 32 US S30, CATALOG 412373 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0824-2023 |
ETEST CLINICAL FLUCONAZOLE FL 256 US S30, CATALOG 412333 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0823-2023 |
ETEST CLINICAL ERYTHROMYCIN EM 256 US S30, CATALOG 412333 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0822-2023 |
ETEST CLINICAL ERTAPENEM ETP 32 US S30, CATALOG 412331 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0821-2023 |
ETEST CLINICAL CEFTRIAXONE TX 32 US S30, CATALOG 412302 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0820-2023 |
ETEST CLINICAL BENZYLPENICIL PG 256 US S30, CATALOG 412262 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0819-2023 |
ETEST CLINICAL AMOXI/CLAV 2/1 XL 256 US S30, CATALOG 412252 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0818-2023 |
ETEST CLINICAL AMOXI/CLAV 2/1 XL 256 US S30, CATALOG 412240 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0817-2023 |
ETEST CLINICAL AMIKACIN AK 256 US S30, CATALOG 412218 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0816-2023 |
VITEK MS CLINICAL REAGENT VITEK MS-FA, CATALOG 411072 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0815-2023 |
VITEK MS CLINICAL REAGENT VITEK MS-CHCA, CATALOG 411071 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0814-2023 |
VITEK MS CLINICAL REAGENT VITEK MS-DS, CATALOG 410893 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0813-2023 |
PPM INDUSTRY COMBOURG LPT BROTH (4X3L), CATALOG 410849 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0812-2023 |
ASTUTE REAGENT ASTUTE ELECTRONIC QUALITY CTRL DEVICE US, CATALOG 400016 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0811-2023 |
TEMPO REAGENTS TEMPO CONTROL KIT , CATALOG 80000 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0810-2023 |
API JAMES 2AMP, CATALOG 70542 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0809-2023 |
API NIT1 NIT2 REAGENTS, CATALOG 70442 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0808-2023 |
API VP1 VP2 REAGENTS, CATALOG 70422 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0807-2023 |
BIOBALL P.AERUGINOSA NCTC 12924 550X20, CATALOG 56017 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0806-2023 |
BIOBALL B.SPIZIZENII NCTC 10400 550X20, CATALOG 56012 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0805-2023 |
BIOBALL A.BRASILIENSIS NCPF 2275 550X20, CATALOG 56011 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0804-2023 |
PPM CLINICAL CRAPONNE CHROMID C.DIFFICILE 20 PLT, CATALOG 43871 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0803-2023 |
PPM CLINICAL CRAPONNE CHROMID VRE 20 PLT US, CATALOG 43851 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0802-2023 |
PPM CLINICAL CHROMID MRSA 20 PLT US, CATALOG 43841 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0801-2023 |
PPM CLINICAL CRAPONNE CLINICAL CHROMID CANDIDA 20 PLATES, CATALOG 43631 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0800-2023 |
VIDAS CLINICAL QCV-QUALITY CONTROL VIDAS 60T, CATALOG 30706 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0799-2023 |
VIDAS CLINICAL VIDAS SALMONELLA 60T, CATALOG 30702 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0798-2023 |
VIDAS CLINICAL VIDAS LISTERIA 60T, CATALOG 30700 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0797-2023 |
VIDAS CLINICAL VIDAS RUB IGG 60 TESTS, CATALOG 30226 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0796-2023 |
VIDAS CLINICAL VIDAS MEASLES IGG 60TESTS, CATALOG 30219 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0795-2023 |
VIDAS CLINICALVIDAS VARICEL. ZOSTER IGG 60 T, CATALOG 30217 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0794-2023 |
PREVI COLOR GRAM REAGENT NOZZLE CLEANING SOLUTION (NA), CATALOG 29588 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0793-2023 |
PREVI COLOR GRAM REAGENT IODINE-B (NA), CATALOG 29586 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0792-2023 |
VITEK 2 REAGENT AST-GP67 TEST KIT 20 CARDS, CATALOG 22226 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0791-2023 |
VITEK 2 REAGENT NH TEST KIT VTK2 20 CARDS, CATALOG 21346 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0790-2023 |
VITEK 2 REAGENT BCL TEST KIT VTK2 20 CARDS, CATALOG 21345 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0789-2023 |
VITEK 2 REAGENT YST TEST KIT VTK2 20 CARDS, CATALOG 21343 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0788-2023 |
VITEK 2 REAGENT GP TEST KIT VTK2 20 CARDS, CATALOG 21342 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0787-2023 |
VITEK 2 REAGENT GN TEST KIT VTK2 20 CARDS, CATALOG 21341 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0786-2023 |
API STAPH 25 STRIPS+25 MEDIA, CATALOG 20500 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0785-2023 |
API 20 C AUX 25 STRIPS+25 MEDIA, CATALOG 20210 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0784-2023 |
API 20 E 25 STRIPS, CATALOG 20100 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-11-09 |
Class II
|
Ongoing |
Z-0783-2023 |
API LISTERIA 10 STRIPS+10 MEDIA, CATALOG 10300 |
— |
As the temperature and time out of range was exceeded, product performances cannot be guaranteed |
| 2022-07-20 |
Class II
|
Ongoing |
Z-1564-2022 |
NUCLISENS easyMAG Magnetic Silica, REF 280133, 24 x 1.2mL bottles/kit. |
— |
There is a potential of no result leading to possible delayed result due to contamination of extraction reagents. |
| 2022-07-13 |
Class II
|
Ongoing |
Z-1640-2022 |
VITEK 2 / MYLA VITEK 2 Software versions 8.01, 8.02 (veterinary use only), 9.01, 9.02, 9.03 MYLA Software versions V… |
— |
There are 7 reported software anomalies that may affect use of the device. |
| 2022-03-29 |
Class II
|
Ongoing |
Z-1059-2022 |
VITEK 2 Systems and VITEK 2 with MYLA. |
— |
Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected int… |
| 2022-02-09 |
Class II
|
Terminated |
Z-0742-2022 |
VIDAS CMV IgM, 30 tests, REF 30205-01, IVD, Rx. |
— |
Invalid calibrations (out of range high) while using the product. |
| 2021-12-20 |
Class II
|
Terminated |
Z-0525-2022 |
API ID 32 C Strips, Reference 32200, Rx Only, IVD API ID 32 C is a qualitative standardized system for the identifica… |
— |
Risk of misidentification or delayed results with the absence of identification due to thermoforming defect. |
| 2021-11-09 |
Class II
|
Terminated |
Z-0422-2022 |
RAPID 20 E, REF 20701, System for the identification of Enterobacteriaceae in 4 hours, 25 strips/package, IVD. |
— |
There is an error in the new version of the package insert. The Reading Table lists the incorrect color for a negative result for the CIT test. |
| 2021-09-30 |
Class II
|
Terminated |
Z-0243-2022 |
API 50 CH, Model 50300 |
— |
There is potential to provide an incorrect organism identification. |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0439-2022 |
VIDAS SARS-COV-2 IgG, Ref 423834-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0438-2022 |
VIDAS SARS-COV-2 IgM, Ref 423833-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0437-2022 |
VIDAS Lyme IgG II (LYM), REF 417401 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0436-2022 |
VIDAS Lyme IgM II (LYM), REF 416436 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0435-2022 |
VIDAS Estradiol II, REF 30431-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0434-2022 |
VIDAS FSH, REF 30407-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0433-2022 |
VIDAS LH, REF 30406-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0432-2022 |
VIDAS RUB IgG (RBG), REF 30226 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0431-2022 |
VIDAS TOXO IgG Avidity (TXGA), REF 30222-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0430-2022 |
VIDAS Measles IgG (MSG), REF 30219 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0429-2022 |
VIDAS Mumps IgG (MPG), REF 30218 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0428-2022 |
VIDAS B.R.A.H.M.S. PROCALCITONIN PCT, REF 30450-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0427-2022 |
VIDAS DEX@, Dimer Exclusion II, REF 30455-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0426-2022 |
VIDAS Clostridium Difficile GDH, REF 30125-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0425-2022 |
VIDAS Clostridium Difficile Toxin A&B, CDAB,REF 30118-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-09-22 |
Class II
|
Ongoing |
Z-0424-2022 |
VIDAS PRG Progesterone, REF 30409-01 |
— |
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to pote… |
| 2021-06-24 |
Class II
|
Terminated |
Z-2226-2021 |
ETEST IMIPENEM RELEBACTAM. in vitro diagnostic |
— |
Major errors (Resistant result instead of Susceptible result) were observed with Pseudomonas aeruginosa strains. |
| 2021-03-16 |
Class II
|
Terminated |
Z-1408-2021 |
VIDAS¿ CMV IgM 30 Tests |
— |
bioMerieux received complaints about calibration issue observed on several lots of VIDAS¿ CMV IgM (ref 30205 ; 30205-01). |
| 2021-01-28 |
Class II
|
Terminated |
Z-1308-2021 |
VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to H… |
— |
bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Co… |
| 2021-01-28 |
Class II
|
Terminated |
Z-1307-2021 |
VITEK 2 (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (syst… |
— |
bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Co… |
| 2020-03-11 |
Class II
|
Ongoing |
Z-1742-2020 |
The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as a… |
— |
There is a potential for the occurrence of false positive OXSF results for some Staphylococcus aureus strains in conjunction with the VITEK 2 AST-P652 test card… |
| 2020-02-05 |
Class II
|
Terminated |
Z-1534-2020 |
VITEK 2 card AST-N371 - Product Usage: The VITEK 2 is an automated system consisting of instruments, software, and reage… |
— |
The top seal of some of the pouches was compromised which can allow moisture to enter that can impact some antibiotics on the card. |
| 2020-01-24 |
Class II
|
Terminated |
Z-1793-2020 |
PREVI Isola Insrument and PREVI Isola Printer spares PSUs (power supply units), P1028888-001 PSU (used in Zebra LP2824 … |
— |
The recall has been initiated due to potential for overheating or a fire hazard associated with the Zebra LP2824 Plus printer PSU (power supply unit) located w… |