| 2026-05-06 |
Class I
|
Ongoing |
Z-2278-2026 |
Design Options Spinal Tray Spinal Needle Anesthesia procedure kits: Material Description (Material Number): SSK H M C… |
— |
These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huon Co, Ltd., has issued a voluntary re… |
| 2026-05-06 |
Class I
|
Ongoing |
Z-2277-2026 |
Spinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. (90 mm), B… |
— |
These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huon Co, Ltd., has issued a voluntary re… |
| 2026-05-06 |
Class I
|
Ongoing |
Z-2276-2026 |
Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREP (333867… |
— |
These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huon Co, Ltd., has issued a voluntary re… |
| 2026-03-19 |
Class I
|
Ongoing |
Z-1803-2026 |
B Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 4.8MM PUMP SEGMENT. Model Number: B3-4630M4705. |
— |
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure. |
| 2026-03-19 |
Class I
|
Ongoing |
Z-1802-2026 |
B. Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 2.6MM PUMP SEGMENT. Model Number: B3-3632M3705. |
— |
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure. |
| 2026-03-19 |
Class I
|
Ongoing |
Z-1801-2026 |
B. Barun Hemodialysis Bloodlines, StreamLine Bloodline Long Version, FMC. Model Number: SL-2000M2095L. |
— |
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure. |
| 2026-03-19 |
Class I
|
Ongoing |
Z-1800-2026 |
B. Braun Hemodialysis Bloodlines, STREAMLINE FRESENIUS FOR DAVITA. Model Number: SL-2000M2095DA. |
— |
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure. |
| 2026-03-19 |
Class I
|
Ongoing |
Z-1799-2026 |
B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096 |
— |
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure. |
| 2026-03-19 |
Class I
|
Ongoing |
Z-1798-2026 |
B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG DR. Model Number: SL-2010M2096A. |
— |
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure. |
| 2026-02-02 |
Class II
|
Ongoing |
Z-1483-2026 |
Brand Name: B. Braun Product Name: 21GA WINGED INF SET Model/Catalog Number: 7M2802 Software Version: N/A Product De… |
— |
The potential for the needle tip to be dull/blunt, difficult to advance, and/or break. |
| 2026-02-02 |
Class II
|
Ongoing |
Z-1482-2026 |
Brand Name: B. Braun Product Name: 21GA WINGED INFUSION, SINGLE PK Model/Catalog Number: 7A3842 Software Version: N/A… |
— |
The potential for the needle tip to be dull/blunt, difficult to advance, and/or break. |
| 2026-02-02 |
Class II
|
Ongoing |
Z-1481-2026 |
Brand Name: B. Braun Product Name: BBraun Medical 21G X4.4CM Winged INF Model/Catalog Number: 7B3050 Software Version… |
— |
The potential for the needle tip to be dull/blunt, difficult to advance, and/or break. |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0718-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0717-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0716-2026 |
Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0715-2026 |
Vista set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space L… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0714-2026 |
ULTRASITE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0713-2026 |
SAFELINE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spac… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0712-2026 |
SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0711-2026 |
SafeDAY set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0710-2026 |
RATE FLOW set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0709-2026 |
Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0708-2026 |
Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0707-2026 |
Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0706-2026 |
Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0705-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0704-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0703-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0702-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0701-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0700-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0699-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0698-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0697-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0696-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0695-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0694-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0693-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0692-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0691-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0690-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0689-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0688-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0687-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0686-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0685-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0684-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0683-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0682-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0681-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0680-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0679-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0678-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0677-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0676-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0675-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0674-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0673-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0672-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0671-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0670-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0669-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0668-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0667-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0666-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0665-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0664-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0663-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0662-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0661-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0660-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0659-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0658-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0657-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0656-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0655-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0654-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0653-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0652-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0651-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0650-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0649-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0648-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0647-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0646-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0645-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0644-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0643-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0642-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0641-2026 |
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infus… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0640-2026 |
Infusomat utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space L… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0639-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0638-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0637-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0636-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0635-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0634-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0633-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0632-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0631-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0630-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0629-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0628-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0627-2026 |
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Spa… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0626-2026 |
Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space La… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0625-2026 |
Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space La… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0624-2026 |
Caresite utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space La… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0623-2026 |
CARESAFE utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space La… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0622-2026 |
Burette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0621-2026 |
Burette Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0620-2026 |
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's In… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0619-2026 |
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's In… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0618-2026 |
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's In… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0617-2026 |
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's In… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0616-2026 |
Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's In… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0615-2026 |
B.BRAUN utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Lar… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0614-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0613-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0612-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0611-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0610-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0609-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0608-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0607-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0606-2026 |
Anesthesia IV Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0605-2026 |
ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Lar… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0604-2026 |
ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Lar… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-10-29 |
Class II
|
Ongoing |
Z-0603-2026 |
ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Lar… |
— |
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion). |
| 2025-09-30 |
Class I
|
Ongoing |
Z-0070-2026 |
Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M20… |
— |
The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0126-2026 |
Design Options, PERIFIX FX Epidural Anesthesia Tray. Product Code: CESK. Product Descriptions: 1. CESK ROBERT WOOD J… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0125-2026 |
Combined Spinal and Epidural Anesthesia Tray 2, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN 27 Ga.… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0124-2026 |
Design Options, Pain Management Tray. Product Code: SESK. Product Descriptions: 1. SESK PARKLAND MEMORIAL FX CSE TRAY.… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0123-2026 |
Combined Spinal and Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN 27 Ga. x… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0122-2026 |
ESPOCAN Combined Spinal and Epidural Anesthesia Tray, ESPOCAN 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN … |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0121-2026 |
PERIFIX FX Continuous Epidural Anesthesia Set, PERIFIX Tuohy Epidural Needle - 17 Ga. x 3-1/2 in. (8.9 cm) - Winged, PER… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0120-2026 |
PERIFIX FX Continuous Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 1… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0119-2026 |
PERIFIX FX Continuous Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 1… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0118-2026 |
PERIFIX FX Continuous Epidural Anesthesia Tray, 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Fixed Wings, 19 Ga. … |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0117-2026 |
PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged a… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0116-2026 |
PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged a… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0115-2026 |
PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (8.9cm) Tuohy Epidural Needle - Winged an… |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-09-18 |
Class II
|
Ongoing |
Z-0114-2026 |
PERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (9cm) Tuohy Epidural Needle - Winged and … |
— |
Potential for the lid of the catheter connector to be in the incorrect position. |
| 2025-07-09 |
Class I
|
Ongoing |
Z-2176-2025 |
Microbore Extension Set with 0.2 Micron Filter. Catalog Number: V6215. for use with IV tubing. |
— |
Label is incorrect. The product label indicates that the device contains an air eliminating filter. However, the filter utilized on this extension set is not in… |
| 2025-02-17 |
Class II
|
Ongoing |
Z-1397-2025 |
Introcan Safety 3 Closed IV Catheter 22 Ga. x 1 in., PUR, Winged. Passive safety IV catheter. |
— |
Increased risk for valve housing detachment causing leakage. |
| 2024-12-04 |
Class II
|
Ongoing |
Z-1021-2025 |
Design Options COMBINED SPINAL/EPIDURAL TRAY- Intended for administration of analgesic/anesthetic agent to intrathecal … |
— |
Incorrect catheter connector within the kit resulting in inability to provide a secure connection to the epidural catheter. The incorrect catheter connector is… |
| 2024-08-15 |
Class II
|
Ongoing |
Z-3179-2024 |
STREAMLINE BLOODLINE SET FOR DIALOG DR-Tubing to be used to transfer blood between a patient and a haemodialyser during … |
— |
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion… |
| 2024-08-15 |
Class II
|
Ongoing |
Z-3178-2024 |
STREAMLINE BLOODLINE SET FOR DIALOG-Tubing to be used to transfer blood between a patient and a haemodialyser during hae… |
— |
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion… |
| 2024-08-15 |
Class II
|
Ongoing |
Z-3177-2024 |
STREAMLINE LONG-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model… |
— |
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion… |
| 2024-08-15 |
Class II
|
Ongoing |
Z-3176-2024 |
STREAMLINE BLOODLINE LONG VERSION, FMC-Tubing to be used to transfer blood between a patient and a haemodialyser during … |
— |
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion… |
| 2024-08-15 |
Class II
|
Ongoing |
Z-3175-2024 |
STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodi… |
— |
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion… |
| 2024-08-15 |
Class II
|
Ongoing |
Z-3174-2024 |
STREAMLINE FRESENIUS FOR DAVITA-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodi… |
— |
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion… |
| 2024-08-15 |
Class II
|
Ongoing |
Z-3173-2024 |
SL BLOODLINE FOR FMC 2008 SERIES-Tubing to be used to transfer blood between a patient and a haemodialyser during haemod… |
— |
Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion … |
| 2024-08-08 |
Class II
|
Ongoing |
Z-3144-2024 |
AET36 ANESTH EXT SET - Anesthesia Extension Set - Normally Closed Dual Manifold, Four-way Stopcock, Model/Catalog Number… |
— |
Potential for product leakage from the Spin Lock Connector on the distal end of the device due to the migration of excess solvent during the assembly process. |
| 2024-08-07 |
Class I
|
Ongoing |
Z-2972-2024 |
Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U |
— |
Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge d… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2404-2024 |
OUTLOOK PUMP SET 3 CARESITE LADS 133 IN.-Used with an electrically-powered infusion pump for intravenous (IV) administra… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2403-2024 |
Infusomat UNIV. 15 DROP PUMP SET, 3 ULTRASITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV)… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2402-2024 |
Outlook OUTLOOK PUMP SET,3 CARESITE LADS,115 IN. Used with an electrically-powered infusion pump for intravenous (IV) ad… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2401-2024 |
Infusomat UNIV 60 DROP PUMP SET, W/2 CARESITE INJ. -Used with an electrically-powered infusion pump for intravenous (IV)… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2400-2024 |
Infusomat SPACE PUMP IV SET W/2 CARESITE, CKVLV -Used with an electrically-powered infusion pump for intravenous (IV) … |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2399-2024 |
Infusomat UNIV. 15 DROP PUMP SET W/2 CARESITE CV- Used with an electrically-powered infusion pump for intravenous (IV) a… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2398-2024 |
Infusomat 60 DROP METRISET PUMP SET, 3 CARESITES Used with an electrically-powered infusion pump for intravenous (IV)… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2397-2024 |
Infusomat UNIV. 15 DROP PUMP SET W 3 CARESITE L.L. Used with an electrically-powered infusion pump for intravenous (IV… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2396-2024 |
Infusomat SPACE PUMP IV SET 15D, 90 IN. Used with an electrically-powered infusion pump for intravenous (IV) administra… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2395-2024 |
Infusomat SPACE PUMP IV SET, 127 IN.-Used with an electrically-powered infusion pump for intravenous (IV) administratio… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2394-2024 |
Infusomat SPACE PUMP IV SET 15D, 110 IN. -Used with an electrically-powered infusion pump for intravenous (IV) administ… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2393-2024 |
Infusomat UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) a… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2392-2024 |
Infusomat UNIV. 15 DROP PUMP SET, 3 CARESITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) a… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2391-2024 |
Infusomat UNIV. 15 DR PUMP SET, 1.2FIL, 2CRSTE ASV-Used with an electrically-powered infusion pump for intravenous (IV)… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2390-2024 |
Infusomat UNIV. 60 DROP PUMP SET, 2 CARESITES, ASV-Used with an electrically-powered infusion pump for intravenous (IV) … |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2389-2024 |
Infusomat 60DROP METRISET PUMP, 3 CARESITES, ASV-Used with an electrically-powered infusion pump for intravenous (IV) a… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2388-2024 |
Infusomat UNIV. 15 DROP PUMP SET, 2 CARESITE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) … |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2387-2024 |
Infusomat SPACE PUMP IV SET, 2 CARESITES, ASV-Used with an electrically-powered infusion pump for intravenous (IV) admin… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2386-2024 |
Infusomat UNIV. 15 DROP PUMP SET W/3 SAFELINE, ASV-Used with an electrically-powered infusion pump for intravenous (IV) … |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2385-2024 |
Infusomat 60 DROP METRISET PUMP SET,3 SFLINE ASV-Used with an electrically-powered infusion pump for intravenous (IV) ad… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2384-2024 |
Infusomat SPACE PUMP IV SET , 120 IN.- Used with an electrically-powered infusion pump for intravenous (IV) administrati… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2383-2024 |
Infusomat UNIV. 15 DROP PUMP SET W/ 3 ULTRASITE LL-Used with an electrically-powered infusion pump for intravenous (IV) … |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2382-2024 |
Infusomat SPACE PUMP SET 15 DROP W/2 SAFEDAY-Used with an electrically-powered infusion pump for intravenous (IV) admini… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2381-2024 |
Infusomat UNIV. 15 DROP PUMP SET W/ 0.2 FILTER- Used with an electrically-powered infusion pump for intravenous (IV) adm… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2380-2024 |
Infusomat 60DROP METRISET PUMP SET, 3 SAFELINE-Used with an electrically-powered infusion pump for intravenous (IV) adm… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2379-2024 |
Infusomat UNIV. 15 DR PUMP SET, 1.2 FIL, 2 CRSITE-Used with an electrically-powered infusion pump for intravenous (IV) a… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2378-2024 |
Infusomat UNIV.15DROP PUMP SET W/3 SAFELINE LL-Used with an electrically-powered infusion pump for intravenous (IV) adm… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2377-2024 |
OUTLOOK IV SET 15DROP W/3 CARESITE-Used with an electrically-powered infusion pump for intravenous (IV) administration o… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-17 |
Class II
|
Ongoing |
Z-2376-2024 |
OUTLOOK IV SET 15DROP W/2 CARESITE-Used with an electrically-powered infusion pump for intravenous (IV) administration o… |
— |
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and th… |
| 2024-06-14 |
Class II
|
Ongoing |
Z-2437-2024 |
Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Non-Wireless Model/Catalog Number: 8713032U Pro… |
— |
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal H… |
| 2024-06-14 |
Class II
|
Ongoing |
Z-2436-2024 |
Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product Description: P… |
— |
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal H… |
| 2024-06-14 |
Class II
|
Ongoing |
Z-2435-2024 |
Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product Description… |
— |
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal H… |
| 2024-06-14 |
Class II
|
Ongoing |
Z-2434-2024 |
Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Softwar… |
— |
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal H… |
| 2024-06-14 |
Class II
|
Ongoing |
Z-2433-2024 |
Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U Product D… |
— |
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal H… |
| 2024-05-17 |
Class II
|
Ongoing |
Z-2134-2024 |
Stimuplex A, 30 DEG, 22GX2", 0.70x50mm, Catalogue Number: 4894502 |
— |
The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP. |
| 2024-05-17 |
Class II
|
Ongoing |
Z-2133-2024 |
Stimuplex A, 30 DEG, 20GX6", 0.90x150mm, Catalogue Number: 4894278 |
— |
The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP. |
| 2024-05-17 |
Class II
|
Ongoing |
Z-2132-2024 |
Stimuplex A, 30 DEG, 21GX4", 0.80x100mm, Catalogue Number: 4894260 |
— |
The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP. |
| 2024-05-17 |
Class II
|
Ongoing |
Z-2131-2024 |
Stimuplex A, 30 DEG, 24GX1", 0.55x25mm, Catalogue Number: 4894251. |
— |
The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP. |
| 2024-01-22 |
Class II
|
Ongoing |
Z-1150-2024 |
Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, 5 mL Glass Luer Lock LOR Tray (Kit) A collection of device… |
— |
Kits should contain Filter Straws with Standard Luer Connections however, the kits contain Filter Straws with Neuraxial connectors preventing interface with the… |
| 2024-01-22 |
Class II
|
Ongoing |
Z-1149-2024 |
Epidural tray. PERIFIX 17 Ga. x 3-1/2 in. (90 mm) Tuohy, PERIFIX FX 19 Ga. Springwound Closed Tip Catheter, 5 mL Glass… |
— |
Kits should contain Filter Straws with Standard Luer Connections however the kits contain Filter Straws with Neuraxial connectors preventing interface with the … |
| 2023-11-01 |
Class I
|
Ongoing |
Z-0355-2024 |
Infusomat Administration Sets as listed below: 362031 Infusomat Space Pump Set with Caresite Injection Site; 3… |
— |
In affected products, the tubing outer diameter may vary in size or the length of tubing may be too short. Both issues may lead to false air-in-line alarms, whi… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0136-2024 |
BULK LARGEBORE MALE LL FITTING, Article No. M4341750SN. IV administration component. |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0135-2024 |
30ML BURETTE ASSY WITH STOPCOCK, BNS, Article No. 7A3970. used in a contrast management system. |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0134-2024 |
ALLERGAN UNIVERSAL FILL KIT, REF No. 7M2804. for tissue expander inflation. |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0133-2024 |
THERASPHERE ADMIN SET, Article No. 7M2650. Canada only. Sterile, single use administration set used to deliver Yttrium-… |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0132-2024 |
TERUMO 60" TUBE SET W/HIGH CRACK CHK VLV, Article No. 7S3070. Singapore only. IV injection set. |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0131-2024 |
30 ML 2PORT BURETTE ASSEMBLY, Article No. 7A3950, used in a contrast management system. |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0130-2024 |
AMO ADMINISTRATION SET, Article No. 7A3817, REF OM2505191, for ophthalmologic use. |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0129-2024 |
SA3000 VENTED MICRO SPIKE ADAPTER, REF No. 418121, for intravenous use. |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-26 |
Class II
|
Ongoing |
Z-0128-2024 |
FT1000S FLUID TRANSFER SET, REF No. 356048. For direct transfer of fluids between large volume parenteral solution conta… |
— |
Devices may leak as a result of a manufacturing issue. Leakage may result in a delay of therapy, increase risk of bloodstream infection (BSI), and may expose bo… |
| 2023-09-21 |
Class I
|
Ongoing |
Z-0124-2024 |
8713052U, Infusomat Large Volume Pump, Non-Wireless BATTERY PACK; Volumetric Infusion Pump System |
— |
Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to… |
| 2023-09-21 |
Class I
|
Ongoing |
Z-0123-2024 |
8713051U, Infusomat Large Volume Pump, Wireless; Volumetric Infusion Pump System |
— |
Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to… |
| 2023-08-07 |
Class II
|
Ongoing |
Z-2644-2023 |
B.Braun NORM-JECT Luer Lock Solo, 10 ml, Sterile-Intended to be used to deliver (inject) or withdraw an accurate metere… |
— |
The sterile blister packaging may be damaged, and sterility may be compromised |
| 2023-08-07 |
Class II
|
Ongoing |
Z-2643-2023 |
B. Braun NORM-JECT Luer Solo, 10 ml, Sterile-Intended to be used to deliver (inject) or withdraw an accurate metered dos… |
— |
The sterile blister packaging may be damaged, and sterility may be compromised |
| 2023-07-31 |
Class II
|
Ongoing |
Z-2627-2023 |
Perifix¿ / Epidural anesthesia set, medicated |
— |
Product may be mislabeled with an incorrect lid stock label. |
| 2023-06-05 |
Class II
|
Ongoing |
Z-2003-2023 |
Design Options¿ / Epidural anesthesia kit (10 count carton) |
— |
Kits were assembled with the incorrect Filter Straw. |
| 2023-06-05 |
Class II
|
Ongoing |
Z-2002-2023 |
Perifix¿ / Epidural anesthesia kit (10 count carton) |
— |
Kits were assembled with the incorrect Filter Straw. |
| 2022-09-15 |
Class II
|
Ongoing |
Z-0092-2023 |
CARESITE SMALLBORE EXT SET T-PORT 5 IN., Product Code 470114, 100 units per carton. The A6 Access Device is a valve int… |
— |
The extension set has a high probability of potential failure which could lead to possible leakage or detachment at the device's bonded joint during use. The is… |
| 2022-09-15 |
Class II
|
Ongoing |
Z-0091-2023 |
Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended … |
— |
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may re… |
| 2022-09-15 |
Class II
|
Ongoing |
Z-0090-2023 |
ULTRASITE US04TL Smallbore T-Port Extension Set, Product Code 474515, 100 units per carton. to provide needle-free acce… |
— |
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may r… |
| 2022-09-01 |
Class II
|
Ongoing |
Z-0009-2023 |
Introcan Safety FEP 14G, 2.2x32MM-US, Product Code 4251890-02. A passive anti-needle device to provide venous or arteri… |
— |
Potential for leakage at the catheter hub. |
| 2022-02-07 |
Class II
|
Terminated |
Z-0819-2022 |
PERFUSOR SPACE - US VERSION. Intended Use: syringe infusion pump that can be used to configure customized, tailor-made s… |
— |
Units not properly inspected prior to release to the market, may not function as indicated in the product labeling, resulting in syringe detection and infusion … |
| 2022-01-31 |
Class II
|
Terminated |
Z-0727-2022 |
Dispensing Pins for Air Inlet Filter, Product Catalog No. 412005, DP2500 Dispensing Pin W/Valve. A valve for aspiratio… |
— |
This lot may have an elevated risk of leakage from the Air-Inlet Filter of the device. Leakage of the Air-Inlet Filter may increase the risk of decreasing the d… |
| 2022-01-28 |
Class II
|
Terminated |
Z-0873-2022 |
Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (END) - VERSION 2, Articl… |
— |
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in… |
| 2022-01-28 |
Class II
|
Terminated |
Z-0872-2022 |
Conductivity Sensors utilized on Dialog and Dialog+ Hemodialysis Systems. Conductivity Sensor (BIC) - VERSION 2, Articl… |
— |
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2560-2021 |
Outlook Pump Blood Administration Set with SAFELINE Injection Site-single use, for use with the Outlook Safety Infusion … |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2559-2021 |
Outlook Pump Primary Administration Set w/ micro Tubing for Epidural Administration-single use, for use with the Outlook… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2558-2021 |
Outlook Pump Add-On Set-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2557-2021 |
OUTLOOK PUMP SET 3 CARESITE LADS 133 IN.-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT … |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2556-2021 |
OUTLOOK PUMP SET, 3 CARESITE LADS, 115 IN-single use, for use with the Outlook Safety Infusion System and/or Horizon NXT… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosin… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2555-2021 |
PUMP SET, 15 DR, 3 CARESITE, 127 in.single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump f… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosin… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2554-2021 |
OUTLOOK PUMP IV SET, 134 IN.-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infu… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2553-2021 |
Y-Type Blood, Outlook Pump Set, Caresite-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT P… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2552-2021 |
OUTLOOK PUMP IV SET W/UNIVERSAL SPIKE-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosin… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2551-2021 |
V1500HP OUTLOOK PUMP SET 118-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infu… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2550-2021 |
ET82HP OUTLOOK PUMP EXTENSION SET-single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2549-2021 |
Outlook Pump Measured Volume Solution Set/ w/ Auto Shut-Off. Single-use for use with the Outlook Safety Infusion System … |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2548-2021 |
10 drops/ml, Priming volume: 46ml, Length: 130 in.-single-use for use with the Outlook Safety Infusion System and/or Hor… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2547-2021 |
60 drops/ml, Priming volume: 25 ml, Length: 120 in.in-single-use for use with the Outlook Safety Infusion System and/or … |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosin… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2546-2021 |
Outlook¿ Pump 15 Drop 0.2 Micron Filtered IV Set with 2 CARESITE¿ Y-Sites: single-use for use with the Outlook Safety In… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2545-2021 |
Outlook Pump Sets w/3 CARESITE Needle-Free Injection Sites: single-use for use with the Outlook Safety Infusion System a… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-08-10 |
Class II
|
Terminated |
Z-2544-2021 |
15 drops/ml, Priming Volume: 27ml, Length: 129 in. Single-use for use with the Outlook Safety Infusion System and/or Hor… |
— |
Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing… |
| 2021-06-25 |
Class II
|
Terminated |
Z-2137-2021 |
Infusomat Space Volumetric Pump Administration Set, Product Code 490100 |
— |
There is a potential for the Anti-free flow clip of the administration set to be inverted which may cause unintended free flow if the user fails to follow on sc… |
| 2021-03-15 |
Class II
|
Terminated |
Z-1509-2021 |
Filtered Triple-Leg Extension Set - Intravascular administration set. Product Catalog number: 477034 - Product Usage: Th… |
— |
BBMI has identified through complaints the potential for the extension set to become occluded or unable to prime. |
| 2020-11-23 |
Class II
|
Terminated |
Z-0719-2021 |
APEX Compounding System Control Panel Module, REF: 601238, Product Code AX1000 - Product Usage: intended for intravascul… |
— |
There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the o… |
| 2020-11-23 |
Class II
|
Terminated |
Z-0718-2021 |
APEX Compounding System Control Panel Module, REF: 601237, Product Code AX1000 - Product Usage: intended for intravascul… |
— |
There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the o… |
| 2020-11-23 |
Class II
|
Terminated |
Z-0717-2021 |
APEX Compounding System Control Panel Module, REF: 601224, Product Code AX1000 - Product Usage: intended for intravascul… |
— |
There is the potential for the compounding system to not immediately interrupt compounding and alert user to the presence of an air bubble exceeding 3% of the o… |