| 2026-04-01 |
Class II
|
Ongoing |
Z-2001-2026 |
i-STAT G3+ cartridge; List Number: 03P78-26; |
— |
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cart… |
| 2026-04-01 |
Class II
|
Ongoing |
Z-2000-2026 |
i-STAT EG6+ cartridge; List Number: 03P77-25; |
— |
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cart… |
| 2026-04-01 |
Class II
|
Ongoing |
Z-1999-2026 |
i-STAT EG7+ cartridge; List Number: 03P76-25; |
— |
Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cart… |
| 2025-08-21 |
Class II
|
Ongoing |
Z-2585-2025 |
i-STAT EG6+ cartridge. List Number: 03P77-25. |
— |
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurem… |
| 2025-08-21 |
Class II
|
Ongoing |
Z-2584-2025 |
i-STAT EG7+ cartridge. List Number: 03P76-25. |
— |
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurem… |
| 2025-08-21 |
Class II
|
Ongoing |
Z-2583-2025 |
i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300 |
— |
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurem… |
| 2025-08-21 |
Class II
|
Ongoing |
Z-2582-2025 |
i-STAT CG4+ cartridge (white). List Number: 03P85-25. |
— |
Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2. |
| 2020-12-22 |
Class II
|
Terminated |
Z-0881-2021 |
Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, … |
— |
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for p… |
| 2020-12-22 |
Class II
|
Terminated |
Z-0880-2021 |
Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose… |
— |
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for p… |
| 2020-01-15 |
Class II
|
Terminated |
Z-1263-2020 |
i-STAT CG4+ cartridge (blue), List No. 03P85-50. UDI (01) 00054749002269 - Product Usage: The i-STAT CG4+ cartridge is i… |
— |
The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and … |
| 2020-01-15 |
Class II
|
Terminated |
Z-1262-2020 |
i-STAT CHEM8+ cartridges (blue), List No. 09P31-25. UDI (01) 00054749001910 - Product Usage: The i-STAT CHEM8+ cartridge… |
— |
The I-STAT BLUE CHEM8+ and CG4+ cartridges are not FDA cleared and do not have CLIA waived status. CLIA recognizes two types of laboratory tests: waived and … |
| 2020-01-14 |
Class II
|
Terminated |
Z-1265-2020 |
i-STAT G3+ cartridge (blue), List No. 03P78-50, UDI (01) 00054749002283 - Product Usage: The i-STAT G3+ cartridge may be… |
— |
Marketed without a 510(k). Abbott Point of Care has determined that it does not have sufficient clinical evidence to fully characterize the performance of the… |